UDI Direct Part Marking on Machined Medical Parts: What the Drawing Has to Say机加工医疗器械零件的 UDI 直接标刻:图纸上必须写清楚的事

A device manufacturer in the Netherlands sent us a drawing set last autumn for a reusable stainless instrument tray insert. The drawing carried a UDI requirement, a DataMatrix symbol position, and a note that the code must grade B or better. It said nothing about which process step the mark belonged to, and nothing about what the surface would look like when the mark was made. The first article passed dimensional inspection and failed the code grade, because the mark had been placed on a bead-blasted area. Twelve weeks went into a fix that was one line on the drawing.

UDI direct part marking is where regulatory requirements, marking physics and machining tolerances meet, and the failure mode is almost never the laser. It is the absence of a process position and a surface specification. This page is about what a machined part drawing has to say so that the mark works the first time.

What UDI Direct Part Marking Actually Requires

The rule is narrower than most people assume, and knowing its shape saves a lot of unnecessary marking.

In the United States the requirement sits in 21 CFR 801.45. A device that must carry a UDI on its label must also bear the UDI permanently marked on the device itself when it is intended to be used more than once and to be reprocessed before each use. The regulation is deliberately technology neutral. Section 801.45(c) asks only for automatic identification and data capture, or any alternative technology, that will provide the UDI on demand, and it names no symbology. The issuing agency requirement sits elsewhere, in 21 CFR part 830 subpart C: the UDI itself has to be built to the rules of an FDA accredited issuing agency, and FDA has accredited three of them, GS1, HIBCC and ICCBBA.

There are four exceptions, and they are written into the same section:

  1. Any type of direct marking would interfere with the safety or effectiveness of the device.
  2. The device cannot be directly marked because it is not technologically feasible.
  3. The device is a single-use device that is subjected to additional processing and manufacturing for the purpose of an additional single use.
  4. The device has been previously marked under paragraph (a) of the section.

There is one more requirement in the same section that is easy to miss: a labeler relying on any of these exceptions has to document the basis for that decision in the design and development records.

The direct marking requirement phased in by device class: Class III devices from 24 September 2016, life supporting and life sustaining devices from 24 September 2017, Class II devices from 24 September 2018, and Class I and unclassified devices from 24 September 2022. One thing catches people out even now: FDA states that implantable devices are not required to be directly marked, because they are not reprocessed before each use.

The practical reading for a machinist is this. The parts that need direct marking are the reusable ones that get reprocessed, so surgical instruments, trays, cases, reusable handpieces and similar hardware. The parts that do not are the single-use components that never leave the sterile barrier. If a customer cannot tell you which of the two they are making, that is the first question to settle, because it changes the drawing.

FDA does not mandate GS1 DataMatrix specifically, but GS1 is one of the accredited issuing agencies and GS1 DataMatrix is the carrier the industry has settled on in practice, because it holds the identifier in a small area and survives repeated sterilization. A QR code that only redirects to a web page does not meet the on demand requirement of paragraph (c)(2), and is not the carrier to put on a drawing. The full text of 21 CFR 801.45 is worth reading once at source rather than through a summary, and the FDA recognised consensus standards database covers the marking and sterilization standards a supplier may cite.

Where the Code Goes on the Part, and How Much Room It Needs

Code placement is a machining decision that gets made too late, usually by whoever is standing at the marking station.

A GS1 DataMatrix carrying a GTIN plus a serial number is not small. GS1's own number for exactly this case is 2.5 by 2.5 mm: the GS1 DataMatrix guideline sets a minimum X-dimension of 0.100 mm for laser direct part marking, and cites a target useable area of 2.5 by 2.5 mm carrying a GTIN plus a serial number, with 0.200 mm as the target module size. Our shop rule is to reserve 3 by 3 mm, because at that size a typical 20 to 24 module symbol lands at roughly 0.13 to 0.15 mm per module, clear of the GS1 floor and with margin left for curvature and reflection. Below 2.5 by 2.5 mm we will tell you the grade is at risk before we quote, not after. The quiet zone around the code matters as much as the code: leave a clear margin of at least one module width on all four sides and the grade will improve without changing anything else.

Four placement rules cover most failures we see:

Put the code on a machined or ground surface, not a blasted one. Bead blasting raises a matte texture that scatters the reader's light and drops the grade, sometimes by a full letter. If the part must be blasted, mark it before the blast and accept a lower grade, or mask the code area during blasting.

Avoid the functional faces. A code on a sealing face, a bearing seat or a cutting edge is a defect waiting to be written. The code belongs on a non-functional surface, and the drawing should say which one.

Allow for the curvature. On a cylindrical part the code has to wrap, which distorts the modules at the edges. A flat pad, a milled recess or a slightly flattened band is cheaper than a better camera.

Keep it away from edges and chamfers. A mark that runs over a radius reads inconsistently and, on an instrument that gets polished during refurbishment, wears away first.

Surface treatment plan for a UDI marked machined component
A code placed on a blasted or coated surface reads poorly. The mark should be planned before the surface step, not after it.打在喷砂或涂层表面的码读取质量很差。打标必须排在表面处理之前规划,而不是之后。

The Mark Belongs Before Passivation, Not After

This is the same argument that applies to instrument marking generally, and it is worth repeating because UDI marking makes it a regulatory exposure rather than a cosmetic one.

A laser mark on stainless or titanium locally removes or re-forms the passive oxide layer. That is true of the dark or engraved marks used for a machine readable code, and less true of annealed or cold marks, which are chosen precisely because they disturb the surface less. On a reusable instrument that will be steam sterilized hundreds of times, that area is where corrosion starts. If the mark is made after passivation, nothing restores the layer. If the mark is made before passivation, the acid bath treats the mark with the rest of the part.

So the UDI mark sits in the route before the passivation or anodize step, not at the end of the line. That single sequencing decision does more for mark durability through reprocessing than any choice of laser parameters, and it costs nothing. It does require the marking decision to be made at the process planning stage rather than when the parts are already finished and sitting in a crate.

On anodized aluminum the same principle applies but the geometry reverses. Anodizing is a conversion coating, not a paint film: roughly half the coating thickness grows outward above the original surface and half consumes the aluminium beneath it, so the part changes size as well as colour. Type II coatings run about 8 to 20 um and Type III hardcoat about 25 to 75 um, and either way the coating ends up on top of any mark made before it, which is where a shallow laser mark loses contrast. The practical routes are to mark after anodizing and accept that the mark is a surface feature, or to mask the code area before anodizing and mark into bare metal afterwards. Either way it is a decision that belongs on the drawing, not on the shop floor. Our anodizing guide covers film growth and masking, the process side of the marking itself is in our laser marking service guide, and the general marking trade-offs are in our part marking methods comparison.

Grading: What B or Better Actually Means

A code either reads or it does not, and a grade is what turns that into something you can specify.

This page covers the UDI specific side of marking: which devices the rule reaches, where the code sits in the process route, and what a drawing has to reserve. The wider comparison of marking methods, their depth and their durability, belongs to our part marking methods comparison, and the identification of a part back to a machining lot is its own subject in our CNC part traceability marking guide. This page assumes the code has to survive regulatory scrutiny, not just a shop floor.

Grading for directly marked parts is done under ISO/IEC TR 29158, the direct part mark quality guideline, which incorporates the earlier AIM DPM-1-2006. It scores a symbol on several axes including cell contrast, cell modulation, fixed pattern damage and axial non-uniformity, and returns one of five letter grades: A, B, C, D or F. There is no E, and F is a fail. The grade for the symbol as a whole is the lowest of the individual parameter grades, which is why one weak parameter sinks an otherwise good mark. Grade B is the level we see asked for most often on medical device drawings, and grade C is regularly accepted on small parts. On the materials side, the same roughness and residue limits that decide whether a mark reads are the ones set for medical device machining cleanliness, and the supplier qualification framework behind them is set out in our ISO 13485 CNC machining guide. What matters for a machining drawing is that the grade depends on the surface, not just the mark, so specifying "grade B" without specifying the surface it will be marked on is only half a specification.

The variables a machine shop can control, in the order that matters:

VariableEffect on gradeWhat we do about it
Surface texture at the markBlasted or as-cast surfaces scatter light and lose a grade or moreMark on machined, ground or milled surfaces, or mask the area
Module sizeBelow roughly 0.1 mm the reader struggles on curved partsReserve 3 by 3 mm minimum, 2.5 by 2.5 mm floor
ContrastLaser parameters and material both matterParameter set developed on the actual grade and finish
Curvature and angleWrapped symbols distort at the edgesFlat pad or recess where the geometry allows
Post-mark processingBlasting, polishing and coating all degrade a markSequence the mark before the destructive steps
Cleaning and passivationRestores the passive layer over the markMark before passivation, never after

Using UDI as a Traceability Spine

Once the code is on the part, it is worth connecting it to everything else, because the cost of doing so is close to zero and the value in a complaint investigation is high.

A serialized DataMatrix that maps back to a machining lot gives you a route from a part in a hospital back to the material certificate, the machine, the inspection record and the operator. That is a different thing from a UDI, which identifies the device model; the serial number is what identifies the unit. We build both into the same mark when the drawing calls for it, and the records that sit behind it are described in our CNC part traceability marking guide and our first article inspection guide.

For a medical program the traceability chain is not optional in either direction. ISO 13485 clause 7.5.9, and specifically 7.5.9.2 for implantable devices, requires traceability records to include the components and materials used and the conditions of the work environment where those could cause the device not to meet its specified safety and performance requirements. It also requires the organization to require its distributors or suppliers of distribution services to keep distribution records available for inspection, and to keep the name and address of the consignee of each shipping package. A serialized mark is the practical way to make that chain work at part level rather than at lot level.

Serialized UDI code on a machined medical component before packaging
A serialized code on the part is what connects a unit in the field back to its material lot, machine and inspection record.零件上的序列化码,是把现场使用的单件追回到材料批次、机床与检验记录的那条线。
Planning the marking position on a machined medical component
The code area has to be reserved at the process planning stage, on a machined surface that will not be blasted or coated after marking.打码区域必须在工艺规划阶段就预留出来,落在打标之后不会被喷砂或涂覆的机加工表面上。

What to Put on the Drawing

Five lines, and they are the difference between a first article that passes and a twelve-week loop.

  1. The UDI requirement: which rule applies, and whether direct part marking is required or excepted.
  2. The carrier and content: symbology, and whether the variable data is a serial number, a lot, a date code or a combination.
  3. The location: a nominated face, with a minimum clear area of 3 by 3 mm.
  4. The surface: machined or ground, not blasted, and whether the area must be masked through any surface treatment.
  5. The process position: before passivation or anodize, and what happens to the mark if the part is reworked.

Add the required grade and the person who will verify it, and the specification is complete. A mark that meets a grade on a drawing is a contract term. A mark that "should be readable" is a conversation after the fact. The measurement side of verifying it, and how a mark is checked against the record it points to, are covered in our CNC inspection methods guide. AAMI is the other reference worth having, since its technical information reports cover the sterilization and reprocessing practice that a directly marked device has to survive.

We build the marking step into the process route on any medical part that carries variable data, and we will tell you at the quote stage if the surface you have nominated will not hold the grade you have asked for. Send the drawing and the UDI requirement and you will have a quotation within 24 hours, with the mark position and the process step written into it rather than left to the shop floor. Start a conversation with our engineering team.

FAQ: common questions on UDI marking

Which devices need UDI direct part marking?

Under 21 CFR 801.45, a device that needs a UDI on its label and is intended to be used more than once and reprocessed before each use must also carry the UDI permanently marked on the device. Single-use devices are generally not required to be directly marked, and there are further exceptions where marking would affect safety or effectiveness or is not technologically feasible.

Does the UDI have to be a DataMatrix?

FDA does not mandate a specific symbology. The rule requires automatic identification and data capture using a method that meets the requirements of an FDA accredited issuing agency, and GS1 DataMatrix is the carrier the industry uses most because it packs the identifier into a small area and survives repeated sterilization.

Can laser marking damage the part?

The mark itself is shallow and does not affect structural performance on a properly developed parameter set. What it does affect is the passive oxide layer on stainless and titanium, which is why the mark belongs before passivation in the process route rather than at the end of the line.

How much room does the code need on the drawing?

Reserve a minimum clear square of 3 by 3 mm for a serialized GS1 DataMatrix, with 2.5 by 2.5 mm as the absolute floor, plus a quiet zone of at least one module width around it. Put it on a machined or ground surface rather than a blasted one, and keep it off sealing faces and cutting edges.

去年秋天,一家荷兰器械厂发来一套图纸,是一件可重复使用不锈钢器械托盘的内衬。图纸上有 UDI 要求、有 DataMatrix 码位,还有一句「码必须达到 B 级或更好」。但图纸没说打标属于哪道工序,也没说打标时表面是什么状态。首件通过了尺寸检验,却在码的评级上不合格,因为码被打在了喷砂面上。为了补上图纸上本来只需要一句话的东西,项目花了十二周。

医疗器械零件的 UDI 直接标刻,是法规要求、打标物理与加工公差三者交汇的地方,而失败的原因几乎从来不是激光,而是「工艺位置」和「表面状态」这两项缺失。本文讲的就是:一张机加工零件图纸需要写清楚什么,才能让标记一次做成。

UDI 直接标刻到底要求什么

这条规则的适用范围比多数人以为的窄,弄清它的形状能省下大量不必要的打标。

在美国,这项要求写在 21 CFR 801.45。必须在标签上载有 UDI 的器械,如果预期被多次使用、且每次使用前需要再处理,就必须同时在器械本体上永久标刻 UDI。该条款刻意保持技术中立:801.45(c) 只要求自动识别与数据采集(或任何替代技术)能够在需要时提供该器械的 UDI,并未点名任何码制。发码机构的要求在别处——21 CFR 第 830 部分 C 子部分:UDI 本身必须按 FDA 认可发码机构的规则生成,FDA 目前认可三家:GS1、HIBCC 与 ICCBBA。

同一节里还写了四种豁免情形:

  1. 任何类型的直接标刻都会影响器械的安全性或有效性。
  2. 该器械无法被直接标刻,因为技术上不可行。
  3. 该器械是一次性使用器械,且经过额外的加工与制造,以供再一次单次使用。
  4. 该器械此前已按本条 (a) 款标刻过 UDI。

同一节还有一条容易漏掉的要求:引用上述任一豁免的贴标商,必须把该决定的依据记录在设计开发文档中。

本体标刻要求按器械类别分阶段生效:III 类器械 2016 年 9 月 24 日;生命支持与生命维持器械 2017 年 9 月 24 日;II 类器械 2018 年 9 月 24 日;I 类及未分类器械 2022 年 9 月 24 日。有一处至今仍常被踩坑:FDA 明确说明植入器械不要求本体直接标刻,因为它们在每次使用前不会再处理。

对机加工厂的实际含义是这样的:需要直接标刻的,是那些会被反复再处理的可重复使用件——手术器械、托盘、器械盒、可重复使用的手持件之类;不需要标刻的,是从不离开无菌屏障的一次性件。如果客户说不清自己做的是哪一类,这就是第一个要定下来的问题,因为它决定图纸长什么样。

FDA 并未强制指定 GS1 DataMatrix,但 GS1 是认可发码机构之一,而 GS1 DataMatrix 是行业实际沉淀下来的载体,因为它能在很小的面积里装下标识符,并且挺得住反复灭菌。一个只做网页跳转的二维码不满足 (c)(2) 的「按需提供」要求,不是该画在图纸上的载体。21 CFR 801.45 的完整原文值得直接读一遍,而不是读二手摘要;FDA 认可共识标准数据库则覆盖供应商可能引用的打标与灭菌标准。

码打在零件哪里,需要多大地方

码位是一个太晚才被决定的加工问题——通常是由站在打标工位前的那个人决定的。

一个携带 GTIN 加序列号的 GS1 DataMatrix 并不小。这个场景下 GS1 自己的数字就是 2.5 × 2.5 mm:GS1 DataMatrix 指南对激光直接标刻规定最小 X 尺寸 0.100 mm,其依据正是「可用标刻面积目标 2.5 × 2.5 mm、数据内容为 GTIN 加序列号」,目标模块尺寸 0.200 mm。我们的厂内法则是预留 3 × 3 mm:在这个尺寸下,20–24 模块的符号折合每模块约 0.13–0.15 mm,明显高于 GS1 下限,给曲面与反光留出余量。低于 2.5 × 2.5 mm 的方案,我们会在报价阶段就告诉你评级有风险,而不是出货之后。码周围的静区(quiet zone)和码本身一样重要:四边各留至少一个模块宽度的干净边距,评级就会改善,而其他什么都不用改。

四条位置规则覆盖了我们见过的大部分失败:

把码打在机加工或磨削面上,不要打在喷砂面上。 喷砂产生的哑光纹理会散射识读器的光并拉低评级,有时整整降一级。如果零件必须喷砂,就在喷砂前打标并接受更低评级,或者在喷砂时遮蔽码区。

避开功能面。 打在密封面、轴承位或刃口上的码,就是一处等着被写进不合格报告的缺陷。码属于非功能面,而图纸应当写明是哪一个面。

给曲率留余量。 在圆柱面上,码必须沿曲面展开,边缘的模块会被扭曲。一块小平面、一个铣出的浅凹台,或者一段轻微压平的带,都比换一台更好的相机便宜。

远离边缘与倒角。 跨过圆角的标记识读不稳定,而且在器械返修抛光时最先被磨掉。

Surface treatment plan for a UDI marked machined component
打在喷砂或涂层表面的码读取质量很差。打标必须排在表面处理之前规划,而不是之后。A code placed on a blasted or coated surface reads poorly. The mark should be planned before the surface step, not after it.

打标排在钝化之前,不是之后

这和器械打标的通用道理是同一条,但 UDI 打标把它从外观问题升级成了法规暴露,所以值得再说一遍。

在不锈钢或钛上,激光打标会局部去除或重塑钝化氧化层。这句话对用于机器识读的深色或雕刻标记成立,对退火标或冷打标则弱得多——后两者被采用的理由恰恰是对表面干扰更小。对一件要经历数百次蒸汽灭菌的可重复使用器械,那个区域就是腐蚀的起点。如果打标发生在钝化之后,没有任何工序能把这层膜补回来;如果打标发生在钝化之前,酸槽会把标记与零件其余部位一起处理。

所以 UDI 标记在工艺路线里的位置是钝化或阳极氧化之前,而不是产线末端。这一个排序决定,对标记在反复再处理中的耐久性所起的作用,比任何激光参数的选择都大,而且它不花钱。它唯一的代价是:打标决策必须在工艺规划阶段做出,而不是等零件已经加工完、躺在料筐里的时候。

在阳极氧化铝上,原理不变但几何关系反转。阳极氧化是转化膜而不是涂层:约一半膜厚在原表面之上向外生长,另一半向内消耗铝基体,所以零件既变颜色也变尺寸。Type II 膜厚约 8–20 µm,Type III 硬膜约 25–75 µm;无论哪种,膜最终都盖在先前做好的标记之上,浅层激光标就是在这里失去对比度的。实际可走的路线是:阳极氧化之后打标,接受标记是一个表面特征;或者在阳极氧化前遮蔽码区,之后在裸金属上打标。无论如何,这都是属于图纸的决定,不属于车间。阳极氧化指南讲了膜层生长与遮蔽,打标工艺本身见激光打标服务指南,各类打标方式的取舍见打标方法对比。

评级:B 级或更好究竟意味着什么

一个码要么读得出来,要么读不出来,而评级就是把这件事实变成可以被规定的东西。

本页管的是 UDI 这一侧:规则覆盖哪些器械、码在工艺路线里的位置、以及图纸必须预留什么。打标方法的横向对比、深度与耐久性,属于我们的打标方法对比;把零件追回到加工批次,是CNC 零件可追溯标记指南的题目。本页的前提是:这个码要经得起法规审查,而不只是车间里能扫。

直接标刻件的评级按 ISO/IEC TR 29158《直接零件标记(DPM)质量指南》执行,该标准整合并取代了早期的 AIM DPM-1-2006。它从单元对比度、单元调制、固定图形损伤、轴向非均匀性等多个维度打分,返回 A、B、C、D、F 五级字母等级——没有 E 级,F 为不合格。符号的总等级取各参数中的最低分,所以一个参数拖后腿就会拉垮整个标记。B 级是我们在医疗器械图纸上见到最多的要求,C 级在小零件上经常被接受。对一张加工图纸而言,要紧的是:评级取决于表面,而不只取决于标记本身,所以只写「B 级」而不写它会被打在哪一种表面上,只算写了一半。

机加工厂能控制的变量,按重要程度排序如下:

变量对评级的影响我们的做法
打标处的表面纹理喷砂面或毛坯面会散射光,掉一级甚至更多打在机加工、磨削或铣削面上,或遮蔽该区域
模块尺寸低于约 0.1 mm 时,识读器在曲面件上会吃力至少预留 3 × 3 mm,下限 2.5 × 2.5 mm
对比度激光参数与材料都有影响在真实牌号与真实表面状态上开发参数组
曲率与角度曲面展开的符号在边缘会失真几何允许时做小平面或凹台
打标后的工序喷砂、抛光、涂覆都会削弱标记把打标排在破坏性工序之前
清洗与钝化会在标记之上重建钝化层打标排在钝化之前,绝不排在其后

材料侧同样如此:决定码能否读出来的那些粗糙度与残留限值,就是医疗器械加工清洁度里设定的那些,而它们背后的供应商准入框架见我们的 ISO 13485 CNC 加工指南。

把 UDI 当作追溯主线来用

码打上零件之后,值得把它接到其他所有记录上,因为这么做的成本接近于零,而在客诉调查中的价值很高。

一个序列化的 DataMatrix,如果能映射回加工批次,就等于给出一条从医院里那件器械追到材质证明、机床、检验记录与操作者的路径。这与 UDI 不是一回事:UDI 标识的是器械型号,序列号标识的是单件。图纸要求时,我们把两者做进同一个标记,其背后的记录体系见我们的CNC 零件可追溯标记指南和首件检验指南。

对医疗项目来说,追溯链在哪个方向都不是可选项。ISO 13485 的 7.5.9 条、以及针对植入性器械的 7.5.9.2 条要求:在这些因素可能导致器械不满足其规定的安全和性能要求时,可追溯性记录应包括所用元器件、材料以及工作环境条件;组织还应要求其经销商或分销服务供方保留分销记录并在检查时可获得,并保留每个货运包装收件人的姓名与地址。序列化标记,就是让这条链落到单件级而不是批次级的实用办法。

Serialized UDI code on a machined medical component before packaging
零件上的序列化码,是把现场使用的单件追回到材料批次、机床与检验记录的那条线。A serialized code on the part is what connects a unit in the field back to its material lot, machine and inspection record.
Planning the marking position on a machined medical component
打码区域必须在工艺规划阶段就预留出来,落在打标之后不会被喷砂或涂覆的机加工表面上。The code area has to be reserved at the process planning stage, on a machined surface that will not be blasted or coated after marking.

图纸上要写哪几行

五行字,就是「首件一次通过」和「十二周返工」之间的差别。

  1. UDI 要求:适用哪条规则,以及是否要求直接标刻或属于豁免。
  2. 载体与内容:码制,以及可变数据是序列号、批次、日期码还是组合。
  3. 位置:指定的某个面,并给出至少 3 × 3 mm 的最小净区域。
  4. 表面:机加工面或磨削面,不得为喷砂面,以及该区域是否必须在表面处理时遮蔽。
  5. 工艺位置:排在钝化或阳极氧化之前,以及零件返工时标记如何处理。

再补上要求的评级等级和由谁负责验证,规格就完整了。图纸上达到某一评级的标记是合同条款;「应该能读」的标记是事后的对话。验证环节的测量方法,以及如何把一个标记与它指向的记录对照核查,见我们的 CNC 检验方法指南。

在任何带可变数据的医疗零件上,我们都会把打标步骤编入工艺路线;如果你指定的表面达不到你要求的评级,我们会在报价阶段就告诉你。把图纸和 UDI 要求发过来,你会在 24 小时内拿到报价,其中写明打码位置与所在工序,而不是留给车间去猜。联系我们的工程团队。

常见问题:UDI 打标

哪些器械需要 UDI 直接标刻?

按 21 CFR 801.45,需要在标签上载有 UDI、且预期多次使用并在每次使用前再处理的器械,必须同时在器械本体上永久标刻 UDI。一次性使用器械通常不要求直接标刻;此外,如果标刻会影响安全性或有效性、或技术上不可行,也有相应豁免。

UDI 必须是 DataMatrix 吗?

FDA 并未强制指定码制。规则要求采用符合 FDA 认可发码机构要求的自动识别与数据采集方式,而 GS1 DataMatrix 是行业使用最多的载体,因为它能把标识符塞进很小的面积,并且挺得住反复灭菌。

激光打标会损伤零件吗?

在参数组开发得当的前提下,标记本身很浅,不影响结构性能。它真正影响的是不锈钢与钛表面的钝化氧化层,这正是标记应排在工艺路线中钝化之前、而不是产线末端的原因。

图纸上要给码留多大地方?

序列化的 GS1 DataMatrix 至少预留 3 × 3 mm 的净方形区域,绝对下限 2.5 × 2.5 mm,并在四周留出至少一个模块宽度的静区。放在机加工面或磨削面上,不要放在喷砂面上,并避开密封面与刃口。